Device Specifications

Number of channels 1
Stimulation modes tDCS, tACS, tRNS, Pulse AM-tACS, Analogue Input Mode
Sham mode None, Single-blind or Double-blind
tDCS Parameters
Stimulation current max ±5 mA (increment 1µA)
Stimulation duration 1-1800s (increment 1s)
Fade-in/out 0-60s (increment 1s)
tACS Parameters
Amplitude of stimulation current (peak-to-peak value) max 5 mA (increment 1µA)
Offset max ±3 mA (increment 1µA)
Stimulation duration 1-1800s (increment 1s)
Fade-in/out 0-60s (increment 1s)
Frequency (carrier) 0.01-600Hz (increment 0.01Hz)
Frequency (modulation) 0-300Hz (increment 0.01Hz)
Modulation index 0-100%
Samples per period 256 (for carrier frequency <300Hz and 128 for carrier frequency ≥300Hz)
Pulse Parameters
Stimulation current max ±5 mA (increment 1µA)
Stimulation duration 1-1800s (increment 1s)
Fade-in/out 0-60s (increment 1s)
Frequency 0.01-600Hz (increment 0.01Hz)
Signal duty cycle 1-99% (increment 1%, for freq. <50Hz) 10-90% (increment 1%, for freq. ≥50Hz)
tRNS Parameters
Sample value range ±(200 - 5000)µA (increment 1µA)
Stimulation duration 1-1800s (increment 1s)
Fade-in/out 0-60s (increment 1s)
Noise distribution rectangular, gaussian
Predefined filters LP: 100Hz, 250Hz, 640Hz, OFF HP: 50Hz, 100Hz, 250Hz, OFF
Analogue Input Mode Parameters
Transfer ratio 2mA per V
Current range ±5 mA
Analogue Bandwidth 1kHz (-3dB)

Other Features

User programs Creating predefined stimulation programs (PRESET) Directly controlling the protocol through the application interface (API)  
Accuracy of current setting <5% for current <1mA, <1% for current >1mA
Voltage max. 35V
Interface Touch screen display, control by PC-USB
Digital inputs and outputs 2.digital inputs 3.digital outputs  
Analogue Outputs Main: 0.5V per mA (single ended output) Low level: 2.5mV per mA (differential +,- & GND)
Power supply Built-in rechargeable battery operating time up to 10h
Dimensions 150 x 115 x 32mm

Safety control
Current density control for a defined electrode; Measurement of impedance (modes: continuous monitoring without stimulation, continuous monitoring with stimulation, single impedance check before stimulation)  
Operation of impedance control Possibility to define a threshold for interrupting stimulation: 10kΩ - 50kΩ Possibility to define user alert threshold: 5kΩ - 15kΩ/off
Electrodes sockets touchproof 1.5mm